Manufacturer: Apera Instruments
Measuring parameters include conductivity, TDS (ppm), salinity (ppt), and temperature. One-tap switch.
Quick and easy 1- to 3-point auto. calibration, recognizing conductivity standards of 1413 µS, 12.88 mS, 84 µS, and 111.8 mS.
Last calibration date check available and calibration reminder can be set up.
Auto. lock, auto. power off function are available.
Temperature unit, conductivity cell constant, reference temperature, and temperature compensation coefficient are all customizable. TDS conversion factor can be set from 0.40 to 1.00.
| Cond. | Measuring Range | Conductivity:0 to 1999 mS/cm, divided into six ranges, automatically switches measuring range. (0 to 19.99) μS/cm; (20.0 to 199.9) μS/cm; (200 to 1999) μS/cm; (2.00 to 19.99) mS/cm; (20.0 to 199.9) mS/cm; (200 to 1999) mS/cm; TDS: (0 to 100) g/L; Salinity: (0 to 100) ppt |
| Resolution | 0.01/0.1/1μS/cm 0.01/0.1 mS/cm | |
| Accuracy | ±1.0% F.S.±1 digit | |
| Temperature Compensation Range | (0 to 50)°C (Auto. or Manual) | |
| Cell constant | 0.1/1/10 cm-1 | |
| Temp. | Measuring Range | 0 to 100? |
| Resolution | 0.1? | |
| Accuracy | ±0.5?±1 digit |
| Data Storage | 500 sets |
| Storage Content | Numbering, measuring value and temperature value, data and time |
| Power | 3* AA alkaline batteries |
| Dimension & Weight | Meter: (88*170*33)mm / 313g Case: (360*270*76)mm / 1.6kg |
| Connector | RCA for temperature, 8-pin for conductivity |
| What's in the box? | EC8500 meter, 2301T-S conductivity/temp. electrode, electrode clip, software flash drive, USB cable, 50ml of conductivity standards (1413µS, 12.88mS), screwdriver (for battery replacement), instruction manual, carrying case |
Manufacturer Part Number/SKU: EC8500-EA
Tags: Apera Instruments